Transpalpebral tonometry after penetrating keratoplasty

Abstract

DOI: 10.32364/2311-7729-2020-20-4-175-179

E.V. Karlova1, A.V. Zolotarev1,2, E.S. Milyudin1, A.E. Pershakova1

1T.I. Eroshevskiy Samara Regional Clinical Ophthalmological Hospital, Samara, Russian Federation

2Samara State Medical University, Samara, Russian Federation

Aim: to assess the efficacy and safety of EASYTON transpalpebral intraocular pressure (IOP) tonometer in the early postoperative period after penetrating keratoplasty (PK).

Patients and Methods: 43 patients after unilateral PK were enrolled. IOP measurements with palpation and EASYTON transpalpebral IOP tonometer in the operated eye were performed daily for 7 days. In addition, IOP measurements with palpation, EASYTON, and the Goldmann applanation tonometer (GAT) in the fellow (non-operated) eye were performed.

Results: Bland-Altman plots demonstrated a rather high degree of the similarity of IOP measurements with palpation, GAT, and EASYTON in the non-operated eyes. Therefore, IOP measurement by palpation is a rather informative technique and can be used to compare the accuracy of EASYTON in the operated eyes of the same patients. During the second step, a high degree of the similarity of IOP measurements with palpation and EASYTON was demonstrated in the operated eyes after PK. Mean IOP spread (as demonstrated by three consecutive IOP measurements with EASYTON) was 1.32 mm Hg thus showing a high degree of repeatability. The lack of the nega tive effects of IOP measurements with EASYTON supports its safety.

Conclusion: EASYTON transpalpebral IOP tonometer prevents the subjectivity of IOP measurements after PK and makes tonometry available to the general public. The lack of the contact with the cornea, a rather high accuracy, a high repeatability, and the lack of negative effects allow for using this device after PK.

Keywords: transpalpebral IOP tonometer, intraocular pressure, penetrating keratoplasty, instrumental tonometry, EASYTON, non-contact tonometry.

For citation: Karlova E.V., Zolotarev A.V., Milyudin E.S., Pershakova A.E. Transpalpebral tonometry after penetrating keratoplasty. Russian Journal of Clinical Ophthalmology. 2020;20(4):175–179. DOI: 10.32364/2311-7729-2020-20-4-175-179.

References
1. De Padua Soares Bezerra B., Chan E., Chakrabarti R., Vajpayee R.B. Intraocular pressure measurement after corneal transplantation. Surv Ophthalmol. 2019;64(5):639–646. DOI: 10.1016/j.survophthal.2019.02.011.
2. Yildirim N., Gursoy H., Sahin A. et al. Glaucoma after penetrating keratoplasty: incidence, risk factors, and management. J Ophthalmol. 2011;2011:951294. DOI: 10.1155/2011/951294.
3. Almousa R., Nanavaty M.A., Daya S.M., Lake D.B. Intraocular pressure control and corneal graft survival after implantation of Ahmed valve device in high-risk penetrating keratoplasty. Cornea. 2013;32(8):1099–1104. DOI: 10.1097/ICO.0b013e31828d2a17.
4. Slonimsky A.Y. Glaucoma and penetrating keratoplasty. Glaukoma. 2004;2:45–50 (in Russ.)
5. Lee H.S., Kim M.S. Influential factors on the survival of endothelial cells after penetrating keratoplasty. Eur J Ophthalmol. 2009;19(6):930–935. DOI: 10.1177/112067210901900606.
6. Fabian I.D., Barequet I.S., Skaat A. et al. Intraocular pressure measurements and biomechanical properties of the cornea in eyes after penetrating keratoplasty. Am J Ophthalmol. 2011;151(5):774–781. DOI: 10.1016/j.ajo.2010.11.007.
7. Vámosi P. Intracameral air following pneumatic noncontact tonometry in a recently post-perforating keratoplasty patient. Clin Ophthalmol. 2014;8:763–765. DOI: 10.2147/OPTH.S60576.
8. Tugeeva E.E., Vorontsova T.N. Opportunities of using transpalpebral tonometer TVGD-01 in pediatric ophthalmology. Clinical Ophthalmology. 2013;14(2):61–63 (in Russ.).
9. Filippova O.M. Transpalpebral tonometery: new opportunities of intraocular pressure registration. Glaukoma. 2004;1:35–38 (in Russ.).
10. Avetisov S.E., Erichev V.P., Antonov A.A. Transpalpebral tonometry: comparative evaluation. Glaukoma 2010;9(3):45–48 (in Russ.).
11. Kushnarevich N. Yu., Iomdina E.N., Bessmertny A.M., Kuzin M.N. Estimation of the accuracy and informativeness of measuring intraocular pressure in patients with their contact lenses on by transpalpebral scleral tonometry. Rossiyskiy oftal’mologicheskiy zhurnal. 2020;13(2):23–28 (in Russ.).
12. Glaucoma. National guide. Ed. E.A. Egorov. M.: GEOTAR-Media; 2014 (in Russ.).
13. Georgaras S., Toliou G., Patrsoura E., Panagopoulos A. Measuring intraocular pressure using finger tension and 2 different instruments in the immediate postoperative period after penetrating keratoplasty. ESCRS Virtual Congress 2020; abstract book.
14. Altman D.G, Bland J.M. Measurement in medicine: the analysis of method comparison studies. Statistician. 1983;32:307–317. DOI: 10.2307/2987937.
15. Carkeet A. A Review of the Use of Confidence Intervals for Bland-Altman Limits of Agreement in Optometry and Vision Science. Optom Vis Sci. 2020;97(1):3–8. DOI: 10.1097/OPX.0000000000001465.

License Creative Commons
This work is licensed under a Creative Commons «Attribution» 4.0 License.
Read the article in PDF in Russian


Previous article
Next article
Diagnostic criteria for asthenic accommo...
0
3424
Ocular surface treatment in dry eye pati...
0
2922
All articles

Распечатать